Scan region, axes and scan direction

Every SPM vendor writes the geometry of a scan differently: some store a centre, some a corner, some no range at all and some no scan direction. This section states the single convention pynxtools-spm writes into NeXus, which raw element of each format it comes from, and how strong the evidence for each reading is.

Where a vendor documents the meaning, the vendor is cited. Where no document was found, the meaning was measured on published data and the measurement is shown, so that anyone can repeat it or overturn it — see Challenge these findings.

What you will find here

Question Short answer Section
What does the scan offset point at? The centre of the scanned area, for every flavour The convention
How are scan_start and scan_end obtained? offset ∓ range/2; never offset and offset + range The convention
What do the axis values mean? Pixel centres, step = range / N The convention
Why are there two lists of axes? @axes describes the image, independent_scan_axes the movement of the tip Two kinds of axes
Where is the scan direction? In the sign of independent_scan_axes, e.g. -Y; a bare X means unknown, or scanned both forward and backward Scan direction as a sign
Which raw key does a value come from? Per-flavour tables Raw-file elements per flavour
How certain is all this? Graded per claim, vendor manual down to none Evidence and confidence

Fields the reader writes

NeXus field Group Meaning Derived from
scan_offset_value_x, _y NXspm_scan_region Centre of the scanned area Vendor key, converted to a centre if the vendor stores a corner
scan_range_x, _y NXspm_scan_region Width and height of the area Vendor key, or points × pitch when not stored
scan_start_x, _y NXspm_scan_region Low edge of the area offset - range/2
scan_end_x, _y NXspm_scan_region High edge of the area offset + range/2
scan_angle_x, _y NXspm_scan_region Rotation of the scan frame Vendor key; not applied to the data
scan_points_x, _y NXspm_scan_pattern Pixels per line, number of lines Vendor key
step_size_x, _y NXspm_scan_pattern Pixel pitch range / points
AXISNAME (X, Y) NXdata Position of each pixel centre offset - range/2 + (i + 0.5) × step
@axes NXdata Axis of each data dimension [Y, X], dimension 0 first
independent_scan_axes NXspm_scan_control Scan axes, fastest to slowest; signed with the scan direction, or bare when that is unknown or both passes were scanned Vendor direction key

The convention

Offset is the centre of the scan area

scan_start_n = offset_n - range_n / 2
scan_end_n   = offset_n + range_n / 2

scan_start = offset with scan_end = offset + range is never used. When a format stores a corner instead of a centre, the corner is shifted first, offset = corner + range/2, and the relations above are applied to it. Start and end are always derived, even where a format has a start-like key of its own.

flowchart LR
    A["offset
(centre)"] --- B["range"] B --> C["scan_start = offset - range/2"] B --> D["scan_end = offset + range/2"] C --> E["axis[0] = scan_start + step/2"] D --> F["axis[-1] = scan_end - step/2"]

Axis values are pixel centres

An axis value is the centre of its pixel, not its edge, so pixel i of N sits at offset - range/2 + (i + 0.5) × step with step = range / N. The first and last axis values are therefore half a pixel inside the scan region, and the midpoint of any axis is exactly the scan offset. Placing the coordinate at the pixel centre is this reader's choice; no vendor prescribes it.

Frames of reference

Frame What it is Fields in it
Scanner (piezo) Position within the scanner range, measured from its undeflected centre scan_offset_value_*, scan_start_*, scan_end_*
Stage Coarse position of the head or sample holder Vendor stage keys, e.g. Bruker \Stage X; never combined with the offset
Sample Where a feature physically sits Can be reconstructed only by combining the stage position with the scanner frame

A non-zero scan_angle_* rotates the scan frame against the sample, so X and Y are the fast and the slow axis of the scan frame, not the x and y of the sample. The angle is recorded but never applied to the data.

Image orientation

Images follow the bottom-left convention: row 0 is the bottom row, column 0 the left column, and both axes ascend. An up scan and a down scan of the same area give the same image; only the order in which the lines were recorded differs.

Two kinds of axes

@axes and independent_scan_axes answer different questions and are written independently of each other.

@axes (NXdata) independent_scan_axes (NXspm_scan_control)
Question answered How should the image be displayed? How did the tip move over the sample?
Order Dimension order, [slow, fast] = [Y, X] Fastest to slowest, [±X, ±Y]
Case Upper case Upper case
Sign Never signed Signed with the scan direction where known
Affected by the image flips Yes, it describes the stored image No, it describes the measurement

Scan direction as a sign

Value Meaning
+Y The axis was traveled towards increasing Y
-Y The axis was traveled towards decreasing Y
Y No single direction is asserted, or the direction is unknown

An axis is signed when the entry holds exactly one pass along it and the format records which way that pass ran.

A bare axis covers two different situations, which the file itself tells apart:

  • Both passes are stored. The entry has a forward and a backward NXdata group, so no single direction applies to the fast axis.
  • The format never records it. Bruker .FLT stores one channel and no slow scan direction, so its Y means unknown, not upward.

Raw-file elements per flavour

Each flavour has its own section with the full key table, how its scan direction is decided and what it leaves unrecorded. Click a flavour to open it.

Flavour Extension Offset Range Points Direction Unit Angle
Nanonis .sxm :SCAN_OFFSET: :SCAN_RANGE: :SCAN_PIXELS: :SCAN_DIR: /Z-Controller/Z unit :SCAN_ANGLE:
Bruker NanoScope .spm \X Offset, \Y Offset \Scan Size, \Aspect Ratio \Samps/line, \Lines \Frame direction In the value, e.g. 20000 nm \Rotate Ang.
Bruker SPMLab .FLT OffsetX, OffsetY ScanRangeX, ScanRangeY ResolutionX, ResolutionY ScanDirection Suffix of the value, e.g. 1.0000 µm Rotation
Omicron / RHK .sm4 RHK_Xoffset, RHK_Yoffset not stored RHK_Xsize, RHK_Ysize sign of RHK_Yscale RHK_X/@unit, RHK_Y/@unit RHK_Angle
No raster .dat, .spm.txt — — — — — —

No raster: Nanonis .dat STS and Bruker .spm.txt

Neither flavour rasters an area, so neither writes a 2D scan region or independent_scan_axes.

Flavour What is written instead
Nanonis .dat STS scan_start_bias, scan_end_bias and scan_offset_bias, read directly from Bias Spectroscopy>Sweep Start (V), >Sweep End (V) and Bias>Offset (V)
Bruker .spm.txt A point_forceSCAN group; start, end and range of both halves of the ramp come from the first and last element of /Calc_Ramp_Ex_nm and /Calc_Ramp_Rt_nm

Evidence and confidence

Grades

Grade Definition
Vendor manual A document published by the instrument maker defines the parameter
Empirical test Raw files from the vendor's software where the geometry can be measured, and only one reading fits
Vendor file content Something the vendor's software writes into the file that fits only one reading
Third-party reader An independent open-source reader, e.g. Gwyddion; how the community reads the value, not how the vendor defines it
None No source found

What each claim rests on

Claim Flavour Grade Source
Offset is the centre Bruker .spm Vendor manual NanoScope 6.13 User Guide, p. 60
Offset is the centre Omicron .sm4 Vendor manual + empirical test RHK R9 User Manual, p. 195, and the measurement
Offset is the centre Nanonis .sxm Vendor file content + third-party readers :Scan>Scanfield: in the files; Gwyddion nanonis.c
Offset is the centre Bruker .FLT Empirical test Measurement
Pitch is \|scale\|, range is N × \|scale\| Omicron .sm4 Third-party readers Gwyddion rhk-sm4.c; MATLAB sm4reader
Sign of RHK_Yscale is the slow direction Omicron .sm4 Third-party reader Gwyddion forum
Row order and image flips All image flavours Third-party reader Gwyddion import modules, verified against every test file

How the empirical test works

A small scan taken inside a larger scan of the same area shows up as a patch of the large image. Where the patch is found is a measurement; where each reading of the offset predicts it differs by (range_big - range_small) / 2 per axis. The small image is resampled to the pixel size of the large one, both are line- and plane-levelled, and the patch is located by normalized cross-correlation.

Pairs whose two scans share the same offset are the clearest: a centre reading puts the small scan in the middle of the large one, a corner reading in a corner, and that holds whatever the axis directions are.

In the results below, Measured is where the small scan was found, while Centre predicts and Corner predicts are the positions computed from the raw offsets under each reading. Whichever prediction the measurement matches is the reading the file follows.

The measured results are shown with the flavour they concern: Bruker SPMLab .FLT and Omicron / RHK .sm4.

Data used

Flavour Files Where License
Bruker .FLT PMIS2-C8_ML2_p1_5__040925135420.SIG_HEIGHT_SENSOR_FRW.FLT, PMIS2-C8_ML2_p1_20__040925132340.SIG_HEIGHT_SENSOR_FRW.FLT In this repository, and in AFM.zip of 10.5281/zenodo.18060234 CC BY 4.0
Omicron .sm4 VT231211_A1_0064.sm4, VT231211_A1_0065.sm4, VT231205_A1_0063.sm4, VT231205_A1_0064.sm4 10.5281/zenodo.14268803; the record holds many more files, so take these four by name CC BY 4.0

Challenge these findings

Four of the claims above rest on measurement or on third-party readers rather than on a vendor document. If you can show one of them is wrong, the reader should change.

Claim Evidence that would overturn it
The .FLT offset is the centre A Veeco, ThermoMicroscopes or Bruker Innova document defining OffsetX, or scans of one area that the centre reading misplaces
The .sm4 range is N × \|scale\| The RHK "SM4 Data File Format" document, or a calibration grating measured against the written range
The .sm4 fast axis has no recorded direction A field in the page header, or the RHK format document, that gives it
Nanonis SCAN_OFFSET is the centre The SPECS TCP Protocol Document, if it says otherwise

Please open an issue with the Challenge a scan-region convention template. A vendor document is the strongest evidence; an openly licensed raw file that the current reading gets wrong is just as welcome, because it can be added to the test set.

References

Source What it establishes Link
NanoScope 6.13 User Guide, pp. 60, 85 Bruker .spm offset is the centre position of the scan PDF
Bruker help, Zoom and Offset Buttons The Offset command centres the scan and updates X/Y Offset Page
Bruker help, Scan View Parameters Tips X/Y Offset use the sample as position reference Page
Bruker help, Frame Commands Frame Up and Frame Down restart the scan at the bottom and the top Page
RHK R9 User Manual, pp. 59, 195, 197 XY Offsets of 0 centre the scan area in the scan range Manual
Park Scientific, User's Guide to AutoProbe CP, pp. 4-17 f. SPMLab's ancestor keeps the offset in scanner coordinates, referenced to the undeflected scanner Scan
Nanonis SXM format description The SXM header tags and the data block layout GXSM tracker
nanonis_control Documents the Nanonis scan frame as centre, size and angle Repository
Gwyddion nanonis.c, nanoscope.c, spmlabf.c, rhk-sm4.c Community reading of each format, including the image flips Source tree
spym Reference Python reader for RHK SM4 Repository
MATLAB sm4reader Computes the SM4 scan size as \|scale\| × points File Exchange